Capillary electrophoresis

Capillary electrophoresis
AcronymCE
ClassificationElectrophoresis
AnalytesBiomolecules
Chiral molecules
Other techniques
Relatedgel electrophoresis
Two-dimensional gel electrophoresis
HyphenatedCapillary electrophoresis mass spectrometry

Capillary electrophoresis (CE) is a family of electrokinetic separation methods performed in submillimeter diameter capillaries and in micro- and nanofluidic channels. Very often, CE refers to capillary zone electrophoresis (CZE), but other electrophoretic techniques including capillary gel electrophoresis (CGE), capillary isoelectric focusing (CIEF), capillary isotachophoresis and micellar electrokinetic chromatography (MEKC) belong also to this class of methods.[1] In CE methods, analytes migrate through electrolyte solutions under the influence of an electric field. Analytes can be separated according to ionic mobility and/or partitioning into an alternate phase via non-covalent interactions. Additionally, analytes may be concentrated or "focused" by means of gradients in conductivity and pH.

Instrumentation

[edit]
Figure 1: Diagram of capillary electrophoresis system

The instrumentation needed to perform capillary electrophoresis is relatively simple. A basic schematic of a capillary electrophoresis system is shown in figure 1. The system's main components are a sample vial, source and destination vials, a capillary, electrodes, a high voltage power supply, a detector, and a data output and handling device. The source vial, destination vial and capillary are filled with an electrolyte such as an aqueous buffer solution. To introduce the sample, the capillary inlet is placed into a vial containing the sample. Sample is introduced into the capillary via capillary action, pressure, siphoning, or electrokinetically, and the capillary is then returned to the source vial. The migration of the analytes is initiated by an electric field that is applied between the source and destination vials and is supplied to the electrodes by the high-voltage power supply. In the most common mode of CE, all ions, positive or negative, are pulled through the capillary in the same direction by electroosmotic flow. The analytes separate as they migrate due to their electrophoretic mobility, and are detected near the outlet end of the capillary. The output of the detector is sent to a data output and handling device such as an integrator or computer. The data is then displayed as an electropherogram, which reports detector response as a function of time. Separated chemical compounds appear as peaks with different migration times in an electropherogram.[2] The technique is often attributed to James W. Jorgensen and Krynn DeArman Lukacs, who first demonstrated the capabilities of this technique.[3] Capillary electrophoresis was first combined with mass spectrometry by Richard D. Smith and coworkers, and provides extremely high sensitivity for the analysis of very small sample sizes. Despite the very small sample sizes (typically only a few nanoliters of liquid are introduced into the capillary), high sensitivity and sharp peaks are achieved in part due to injection strategies that result in a concentration of analytes into a narrow zone near the inlet of the capillary. This is achieved in either pressure or electrokinetic injections simply by suspending the sample in a buffer of lower conductivity (e.g. lower salt concentration) than the running buffer. A process called field-amplified sample stacking (a form of isotachophoresis) results in concentration of analyte in a narrow zone at the boundary between the low-conductivity sample and the higher-conductivity running buffer.

To achieve greater sample throughput, instruments with arrays of capillaries are used to analyze many samples simultaneously. Such capillary array electrophoresis (CAE) instruments with 16 or 96 capillaries are used for medium- to high-throughput capillary DNA sequencing, and the inlet ends of the capillaries are arrayed spatially to accept samples directly from SBS-standard footprint 96-well plates. Certain aspects of the instrumentation (such as detection) are necessarily more complex than for a single-capillary system, but the fundamental principles of design and operation are similar to those shown in Figure 1.

Detection

[edit]

Separation by capillary electrophoresis can be detected by several detection devices. The majority of commercial systems use UV or UV-Vis absorbance as their primary mode of detection. In these systems, a section of the capillary itself is used as the detection cell. The use of on-tube detection enables detection of separated analytes with no loss of resolution. In general, capillaries used in capillary electrophoresis are coated with a polymer (frequently polyimide or Teflon) for increased flexibility. The portion of the capillary used for UV detection, however, must be optically transparent. For polyimide-coated capillaries, a segment of the coating is typically burned or scraped off to provide a bare window several millimeters long. This bare section of capillary can break easily, and capillaries with transparent coatings are available to increase the stability of the cell window. The path length of the detection cell in capillary electrophoresis (~ 50 micrometers) is far less than that of a traditional UV cell (~ 1 cm). According to the Beer-Lambert law, the sensitivity of the detector is proportional to the path length of the cell. To improve the sensitivity, the path length can be increased, though this results in a loss of resolution. The capillary tube itself can be expanded at the detection point, creating a "bubble cell" with a longer path length or additional tubing can be added at the detection point as shown in figure 2. Both of these methods, however, will decrease the resolution of the separation.[4] This decrease is almost unnoticeable if a smooth aneurysm is produced in the wall of a capillary by heating and pressurization, as plug flow can be preserved. This invention by Gary Gordon, US Patent 5061361, typically triples the absorbance path length. When used with a UV absorbance detector, the wider cross-section of the analyte in the cell allows for an illuminating beam twice as large, which reduces shot noise by a factor of two. Together these two factors increase the sensitivity of Agilent Technologies's Bubble Cell CE Detector six times over that of one using a straight capillary. This cell and its manufacture are described on page 62 of the June 1995 issue of the Hewlett-Packard Journal.

Figure 2: Techniques for increasing the pathlength of the capillary: a) a bubble cell and b) a z-cell (additional tubing).[2]

Fluorescence detection can also be used in capillary electrophoresis for samples that naturally fluoresce or are chemically modified to contain fluorescent tags. This mode of detection offers high sensitivity and improved selectivity for these samples, but cannot be utilized for samples that do not fluoresce. Numerous labeling strategies are used to create fluorescent derivatives or conjugates of non-fluorescent molecules, including proteins and DNA. The set-up for fluorescence detection in a capillary electrophoresis system can be complicated. The method requires that the light beam be focused on the capillary, which can be difficult for many light sources.[4] Laser-induced fluorescence has been used in CE systems with detection limits as low as 10−18 to 10−21 mol. The sensitivity of the technique is attributed to the high intensity of the incident light and the ability to accurately focus the light on the capillary.[2] Multi-color fluorescence detection can be achieved by including multiple dichroic mirrors and bandpass filters to separate the fluorescence emission amongst multiple detectors (e.g., photomultiplier tubes), or by using a prism or grating to project spectrally resolved fluorescence emission onto a position-sensitive detector such as a CCD array. CE systems with 4- and 5-color LIF detection systems are used routinely for capillary DNA sequencing and genotyping ("DNA fingerprinting") applications.[5][6]

In order to obtain the identity of sample components, capillary electrophoresis can be directly coupled with mass spectrometers or surface-enhanced Raman spectroscopy (SERS). In most systems, the capillary outlet is introduced into an ion source that utilizes electrospray ionization (ESI). The resulting ions are then analyzed by the mass spectrometer. This setup requires volatile buffer solutions, which will affect the range of separation modes that can be employed and the degree of resolution that can be achieved.[4] The measurement and analysis are mostly done with a specialized.

For CE-SERS, capillary electrophoresis eluants can be deposited onto a SERS-active substrate. Analyte retention times can be translated into spatial distance by moving the SERS-active substrate at a constant rate during capillary electrophoresis. This allows the subsequent spectroscopic technique to be applied to specific eluants for identification with high sensitivity. SERS-active substrates can be chosen that do not interfere with the spectrum of the analytes.[7]

Modes of separation

[edit]

The separation of compounds by capillary electrophoresis is dependent on the differential migration of analytes in an applied electric field. The electrophoretic migration velocity () of an analyte toward the electrode of opposite charge is:

The electrophoretic mobility can be determined experimentally from the migration time and the field strength:

where is the distance from the inlet to the detection point, is the time required for the analyte to reach the detection point (migration time), is the applied voltage (field strength), and is the total length of the capillary.[4] Since only charged ions are affected by the electric field, neutral analytes are poorly separated by capillary electrophoresis.

The velocity of migration of an analyte in capillary electrophoresis will also depend upon the rate of electroosmotic flow (EOF) of the buffer solution. In a typical system, the electroosmotic flow is directed toward the negatively charged cathode so that the buffer flows through the capillary from the source vial to the destination vial. Separated by differing electrophoretic mobilities, analytes migrate toward the electrode of opposite charge.[2] As a result, negatively charged analytes are attracted to the positively charged anode, counter to the EOF, while positively charged analytes are attracted to the cathode, in agreement with the EOF as depicted in figure 3.

Figure 3: Diagram of the separation of charged and neutral analytes (A) according to their respective electrophoretic and electroosmotic flow mobilities

The velocity of the electroosmotic flow, can be written as:

where is the electroosmotic mobility, which is defined as:

where is the zeta potential of the capillary wall, and is the relative permittivity of the buffer solution. Experimentally, the electroosmotic mobility can be determined by measuring the retention time of a neutral analyte.[4] The velocity () of an analyte in an electric field can then be defined as:

Since the electroosmotic flow of the buffer solution is generally greater than that of the electrophoretic mobility of the analytes, all analytes are carried along with the buffer solution toward the cathode. Even small, triply charged anions can be redirected to the cathode by the relatively powerful EOF of the buffer solution. Negatively charged analytes are retained longer in the capillary due to their conflicting electrophoretic mobilities.[2] The order of migration seen by the detector is shown in figure 3: small multiply charged cations migrate quickly and small multiply charged anions are retained strongly.[4]

Electroosmotic flow is observed when an electric field is applied to a solution in a capillary that has fixed charges on its interior wall. Charge is accumulated on the inner surface of a capillary when a buffer solution is placed inside the capillary. In a fused-silica capillary, silanol (Si-OH) groups attached to the interior wall of the capillary are ionized to negatively charged silanoate (Si-O) groups at pH values greater than three. The ionization of the capillary wall can be enhanced by first running a basic solution, such as NaOH or KOH through the capillary prior to introducing the buffer solution. Attracted to the negatively charged silanoate groups, the positively charged cations of the buffer solution will form two inner layers of cations (called the diffuse double layer or the electrical double layer) on the capillary wall as shown in figure 4. The first layer is referred to as the fixed layer because it is held tightly to the silanoate groups. The outer layer, called the mobile layer, is farther from the silanoate groups. The mobile cation layer is pulled in the direction of the negatively charged cathode when an electric field is applied. Since these cations are solvated, the bulk buffer solution migrates with the mobile layer, causing the electroosmotic flow of the buffer solution. Other capillaries including Teflon capillaries also exhibit electroosmotic flow. The EOF of these capillaries is probably the result of adsorption of the electrically charged ions of the buffer onto the capillary walls.[2] The rate of EOF is dependent on the field strength and the charge density of the capillary wall. The wall's charge density is proportional to the pH of the buffer solution. The electroosmotic flow will increase with pH until all of the available silanols lining the wall of the capillary are fully ionized.[4]

Figure 4: Depiction of the interior of a fused-silica gel capillary in the presence of a buffer solution.

In certain situations where strong electroosmotic flow toward the cathode is undesirable, the inner surface of the capillary can be coated with polymers, surfactants, or small molecules to reduce electroosmosis to very low levels, restoring the normal direction of migration (anions toward the anode, cations toward the cathode). CE instrumentation typically includes power supplies with reversible polarity, allowing the same instrument to be used in "normal" mode (with EOF and detection near the cathodic end of the capillary) and "reverse" mode (with EOF suppressed or reversed, and detection near the anodic end of the capillary). One of the most common approaches to suppressing EOF, reported by Stellan Hjertén in 1985, is to create a covalently attached layer of linear polyacrylamide.[8] The silica surface of the capillary is first modified with a silane reagent bearing a polymerizable vinyl group (e.g. 3-methacryloxypropyltrimethoxysilane), followed by introduction of acrylamide monomer and a free radical initiator. The acrylamide is polymerized in situ, forming long linear chains, some of which are covalently attached to the wall-bound silane reagent. Numerous other strategies for covalent modification of capillary surfaces exist. Dynamic or adsorbed coatings (which can include polymers or small molecules) are also common.[9] For example, in capillary sequencing of DNA, the sieving polymer (typically polydimethylacrylamide) suppresses electroosmotic flow to very low levels.[10] Besides modulating electroosmotic flow, capillary wall coatings can also serve the purpose of reducing interactions between "sticky" analytes (such as proteins) and the capillary wall. Such wall-analyte interactions, if severe, manifest as reduced peak efficiency, asymmetric (tailing) peaks, or even complete loss of analyte to the capillary wall.

Efficiency and resolution

[edit]

The number of theoretical plates, or separation efficiency, in capillary electrophoresis is given by:

where is the number of theoretical plates, is the apparent mobility in the separation medium and is the diffusion coefficient of the analyte. According to this equation, the efficiency of separation is only limited by diffusion and is proportional to the strength of the electric field, although practical considerations limit the strength of the electric field to several hundred volts per centimeter. Application of very high potentials (>20-30 kV) may lead to arcing or breakdown of the capillary. Further, application of strong electric fields leads to resistive heating (Joule heating) of the buffer in the capillary. At sufficiently high field strengths, this heating is strong enough that radial temperature gradients can develop within the capillary. Since electrophoretic mobility of ions is generally temperature-dependent (due to both temperature-dependent ionization and solvent viscosity effects), a non-uniform temperature profile results in variation of electrophoretic mobility across the capillary, and a loss of resolution. The onset of significant Joule heating can be determined by constructing an "Ohm's Law plot", wherein the current through the capillary is measured as a function of applied potential. At low fields, the current is proportional to the applied potential (Ohm's Law), whereas at higher fields the current deviates from the straight line as heating results in decreased resistance of the buffer. The best resolution is typically obtained at the maximum field strength for which Joule heating is insignificant (i.e. near the boundary between the linear and nonlinear regimes of the Ohm's Law plot). Generally capillaries of smaller inner diameter support use of higher field strengths, due to improved heat dissipation and smaller thermal gradients relative to larger capillaries, but with the drawbacks of lower sensitivity in absorbance detection due to shorter path length, and greater difficulty in introducing buffer and sample into the capillary (small capillaries require greater pressure and/or longer times to force fluids through the capillary).

The efficiency of capillary electrophoresis separations is typically much higher than the efficiency of other separation techniques like HPLC. Unlike HPLC, in capillary electrophoresis there is no mass transfer between phases.[4] In addition, the flow profile in EOF-driven systems is flat, rather than the rounded laminar flow profile characteristic of the pressure-driven flow in chromatography columns as shown in figure 5. As a result, EOF does not significantly contribute to band broadening as in pressure-driven chromatography. Capillary electrophoresis separations can have several hundred thousand theoretical plates.[11]

Figure 5: Flow profiles of laminar and electroosmotic flow.

The resolution () of capillary electrophoresis separations can be written as:

According to this equation, maximum resolution is reached when the electrophoretic and electroosmotic mobilities are similar in magnitude and opposite in sign. In addition, it can be seen that high resolution requires lower velocity and, correspondingly, increased analysis time.[4]

Besides diffusion and Joule heating (discussed above), factors that may decrease the resolution in capillary electrophoresis from the theoretical limits in the above equation include, but are not limited to, the finite widths of the injection plug and detection window; interactions between the analyte and the capillary wall; instrumental non-idealities such as a slight difference in height of the fluid reservoirs leading to siphoning; irregularities in the electric field due to, e.g., imperfectly cut capillary ends; depletion of buffering capacity in the reservoirs; and electrodispersion (when an analyte has higher conductivity than the background electrolyte).[12] Identifying and minimizing the numerous sources of band broadening is key to successful method development in capillary electrophoresis, with the objective of approaching as close as possible to the ideal of diffusion-limited resolution.

Applications

[edit]

Capillary electrophoresis may be used for the simultaneous determination of the ions NH4+,, Na+, K+, Mg2+ and Ca2+ in saliva.[13]

One of the main applications of CE in forensic science is the development of methods for amplification and detection of DNA fragments using polymerase chain reaction (PCR), which has led to rapid and dramatic advances in forensic DNA analysis. DNA separations are carried out using thin CE 50-mm fused silica capillaries filled with a sieving buffer. These capillaries have excellent capabilities to dissipate heat, permitting much higher electric field strengths to be used than slab gel electrophoresis. Therefore separations in capillaries are rapid and efficient. Additionally, the capillaries can be easily refilled and changed for efficient and automated injections. Detection occurs via fluorescence through a window etched in the capillary. Both single-capillary and capillary-array instruments are available with array systems capable of running 16 or more samples simultaneously for increased throughput.[14]

A major use of CE by forensic biologists is typing of STR from biological samples to generate a profile from highly polymorphic genetic markers which differ between individuals. Other emerging uses for CE include the detection of specific mRNA fragments to help identify the biological fluid or tissue origin of a forensic sample.[15]

Another application of CE in forensics is ink analysis, where the analysis of inkjet printing inks is becoming more necessary due to increasingly frequent counterfeiting of documents printed by inkjet printers. The chemical composition of inks provides very important information in cases of fraudulent documents and counterfeit banknotes. Micellar electrophoretic capillary chromatography (MECC) has been developed and applied to the analysis of inks extracted from paper. Due to its high resolving power relative to inks containing several chemically similar substances, differences between inks from the same manufacturer can also be distinguished. This makes it suitable for evaluating the origin of documents based on the chemical composition of inks. It is worth noting that because of the possible compatibility of the same cartridge with different printer models, the differentiation of inks on the basis of their MECC electrophoretic profiles is a more reliable method for the determination of the ink cartridge of origin (its producer and cartridge number) rather than the printer model of origin.[16]

A specialized type of CE, affinity capillary electrophoresis (ACE), utilizes intermolecular binding interactions to understand protein-ligand interactions.[17] Pharmaceutical companies use ACE for a multitude of reasons, with one of the main ones being the association/binding constants for drugs and ligands or drugs and certain vehicle systems like micelles. It is a widely used technique because of its simplicity, rapid results, and low analyte usage.[18] The use of ACE can provide specific details in binding, separation, and detection of analytes and is proven to be highly practical for studies in life sciences. Aptamer-based affinity capillary electrophoresis is utilized for the analysis and modifications of specific affinity reagents. Modified aptamers ideally exhibit and high binding affinity, specificity, and nuclease resistance.[19] Ren et al. incorporated modified nucleotides in aptamers to introduce new confrontational features and high affinity interactions from the hydrophobic and polar interactions between IL-1α and the aptamer.[20] Huang et al. uses ACE to investigate protein-protein interactions using aptamers. A α-thrombin binding aptamer was labeled with 6-carboxyfluorescein for use as a selective fluorescent probe and was studied to elucidate information on binding sites for protein-protein and protein-DNA interactions.[21]

Capillary electrophoresis (CE) has become an important, cost-effective approach to do DNA sequencing that provides high throughput and high accuracy sequencing information. Woolley and Mathies used a CE chip to sequence DNA fragments with 97% accuracy and a speed of 150 bases in 540 seconds.[22] They used a 4-color labeling and detection format to collect fluorescent data. Fluorescence is used to view the concentrations of each part of the nucleic acid sequence, A, T, C and G, and these concentration peaks that are graphed from the detection are used to determine the sequence of the DNA.[22]

References

[edit]
  1. ^ Kemp G (February 1998). "Capillary electrophoresis: a versatile family of analytical techniques". Biotechnology and Applied Biochemistry. 27 (1): 9–17. doi:10.1111/j.1470-8744.1998.tb01369.x. PMID 9477551. S2CID 45334539.
  2. ^ a b c d e f Baker DR (1995). Capillary Electrophoresis. New York: John Wiley & Sons, Inc.Skoog DA, Holler FJ, Crouch SR (2007). Principles of Instrumental Analysis (6th ed.). Belmont, CA: Thomson Brooks/Cole Publishing.
  3. ^ Jorgenson JW, Lukacs KD (July 1981). "Zone electrophoresis in open-tubular glass capillaries". Analytical Chemistry. 53 (8): 1298–1302. doi:10.1021/ac00231a037. ISSN 0003-2700.
  4. ^ a b c d e f g h i Cunico RL, Gooding KM, Wehr T (1998). Basic HPLC and CE of Biomolecules. Bay Bioanalytical Laboratory. ISBN 978-0-9663229-0-3.
  5. ^ Dovichi NJ, Zhang J (December 2000). "How Capillary Electrophoresis Sequenced the Human Genome This Essay is based on a lecture given at the Analytica 2000 conference in Munich (Germany) on the occasion of the Heinrich-Emanuel-Merck Prize presentation" (PDF). Angewandte Chemie. 39 (24): 4463–4468. doi:10.1002/1521-3773(20001215)39:24<4463::aid-anie4463>3.0.co;2-8. PMID 11169637. Retrieved 2014-04-09.
  6. ^ Butler JM, Buel E, Crivellente F, McCord BR (June 2004). "Forensic DNA typing by capillary electrophoresis using the ABI Prism 310 and 3100 genetic analyzers for STR analysis" (PDF). Electrophoresis. 25 (10–11): 1397–412. doi:10.1002/elps.200305822. PMID 15188225. S2CID 31067288.
  7. ^ He L, Natan MJ, Keating CD (November 2000). "Surface-enhanced Raman scattering: a structure-specific detection method for capillary electrophoresis". Analytical Chemistry. 72 (21): 5348–55. doi:10.1021/ac000583v. PMID 11080886.
  8. ^ Hjertén S (1985). "High-performance electrophoresis: elimination of electroosmosis and solute adsorption". Journal of Chromatography (347): 191–198. doi:10.1016/S0021-9673(01)95485-8.
  9. ^ Doherty EA, Meagher RJ, Albarghouthi MN, Barron AE (January 2003). "Microchannel wall coatings for protein separations by capillary and chip electrophoresis". Electrophoresis. 24 (1–2): 34–54. doi:10.1002/elps.200390029. PMID 12652571. S2CID 25998082.
  10. ^ Madabhushi RS (February 1998). "Separation of 4-color DNA sequencing extension products in noncovalently coated capillaries using low viscosity polymer solutions". Electrophoresis. 19 (2): 224–30. doi:10.1002/elps.1150190215. PMID 9548284. S2CID 221736283.
  11. ^ Skoog DA, Holler FJ, Crouch SR (2007). Principles of Instrumental Analysis (6th ed.). Belmont, CA: Thomson Brooks/Cole Publishing.
  12. ^ Lauer HH, Rozing GP (January 2010). "High Performance Capillary Electrophoresis: A primer" (PDF). Germany: Agilent Technologies. Archived from the original (PDF) on April 13, 2014. Retrieved 2014-04-09.
  13. ^ Hauser PC (2016). "Determination of Alkali Ions in Biological and Environmental Samples". In Astrid S, Helmut S, Roland KO S (eds.). The Alkali Metal Ions: Their Role for Life. Metal Ions in Life Sciences. Vol. 16. Springer. pp. 11–25. doi:10.1007/978-3-319-21756-7_2. ISBN 978-3-319-21755-0. PMID 26860298.
  14. ^ McCord BR, Buel E (2013). "Capillary Electrophoresis in Forensic Genetics". Encyclopedia of Forensic Sciences. Waltham: Academic Press. pp. 394–401. doi:10.1016/B978-0-12-382165-2.00050-7. ISBN 978-0-12-382166-9.
  15. ^ van Oorschot RA, Ballantyne KN (2013). "Capillary electrophoresis in forensic biology". Encyclopedia of Forensic Sciences. Waltham: Academic Press. pp. 560–566. doi:10.1016/B978-0-12-382165-2.00242-7. ISBN 978-0-12-382166-9.
  16. ^ Shallan A, Guijt R, Breadmore M (2013). "Capillary Electrophoresis Basic Principles". In Siegel JA, Saukko PJ, Houck MM (eds.). Encyclopedia of Forensic Sciences. Waltham: Academic Press. pp. 549–559. doi:10.1016/B978-0-12-382165-2.00241-5. ISBN 978-0-12-382166-9.
  17. ^ Chu YH, Avila LZ, Gao J, Whitesides GM (November 1995). "Affinity Capillary Electrophoresis". Accounts of Chemical Research. 28 (11): 461–468. doi:10.1021/ar00059a004.
  18. ^ Neubert RH, Schwarz MA, Mrestani Y, Plätzer M, Raith K (November 1999). "Affinity capillary electrophoresis in pharmaceutics". Pharmaceutical Research. 16 (11): 1663–73. PMID 10571270.
  19. ^ Yu F, Zhao Q, Zhang D, Yuan Z, Wang H (January 2019). "Affinity Interactions by Capillary Electrophoresis: Binding, Separation, and Detection". Analytical Chemistry. 91 (1): 372–387. doi:10.1021/acs.analchem.8b04741. PMID 30392351. S2CID 53217680.
  20. ^ Ren X, Gelinas AD, von Carlowitz I, Janjic N, Pyle AM (October 2017). "Structural basis for IL-1α recognition by a modified DNA aptamer that specifically inhibits IL-1α signaling". Nature Communications. 8 (1): 810. Bibcode:2017NatCo...8..810R. doi:10.1038/s41467-017-00864-2. PMC 5634487. PMID 28993621.
  21. ^ Huang CC, Cao Z, Chang HT, Tan W (December 2004). "Protein-protein interaction studies based on molecular aptamers by affinity capillary electrophoresis" (PDF). Analytical Chemistry. 76 (23): 6973–81. doi:10.1021/ac049158i. PMID 15571349.
  22. ^ a b Woolley AT, Mathies RA (October 1995). "Ultra-high-speed DNA sequencing using capillary electrophoresis chips". Analytical Chemistry. 67 (20): 3676–80. doi:10.1021/ac00116a010. PMID 8644919.

Further reading

[edit]
  • Terabe S, Otsuka K, Ichikawa K, Tsuchiya A, Ando T (January 1984). "Electrokinetic separations with micellar solutions and open-tubular capillaries". Analytical Chemistry. 56 (1): 111–3. doi:10.1021/ac00265a031.
  • Foley JP (July 1990). "Optimization of micellar electrokinetic chromatography". Analytical Chemistry. 62 (13): 1302–8. doi:10.1021/ac00265a031.
  • Segura Carretero A, Cruces-Blanco C, Cortacero Ramírez S, Carrasco Pancorbo A, Fernández Gutiérrez A (September 2004). "Application of micellar electrokinetic capillary chromatography to the analysis of uncharged pesticides of environmental impact". Journal of Agricultural and Food Chemistry. 52 (19): 5791–5. doi:10.1021/jf040074k. PMID 15366822.
  • Cavazza A, Corradini C, Lauria A, Nicoletti I, Stancanelli R (August 2000). "Rapid analysis of essential and branched-chain amino acids in nutraceutical products by micellar electrokinetic capillary chromatography". Journal of Agricultural and Food Chemistry. 48 (8): 3324–9. doi:10.1021/jf991368m. hdl:11381/2441649. PMID 10956110.
  • Rodrigues MR, Caramão EB, Arce L, Ríos A, Valcárcel M (July 2002). "Determination of monoterpene hydrocarbons and alcohols in Majorana hortensis Moench by micellar electrokinetic capillary chromatographic". Journal of Agricultural and Food Chemistry. 50 (15): 4215–20. doi:10.1021/jf011667n. PMID 12105948.
[edit]